cobas b 123 Fluid Pack COOX REF 05169992001 200 Product Usage: The cobas b 123 POC system is a fully automated POC syste

FDA device recall Z-2446-2015 · posted 2015-08-21 · Terminated

Recall details

Recalling firm
Roche Diagnostics Operations, Inc.
Reason for recall
Low PO2 results. QC failures of the PO2 parameter, affecting primarily Levels 1 and 2, caused by a calibration issue with the PO2 parameter. This issue may not be detected since QC results can be below mean values, but still within 2 standard deviations (SD) limits. Potential for erroneously low PO2 results in patient samples, especially in blood samples with PO2 values below 50 mmHg
Action taken
Roche sent an Urgent Medical Device Correction letter dated July 14, 2015 and the Fax Back form via UPS ground to the customers list. The letter identified the affected product, problem and actions to be taken. Consignees are instructed to discontinue use of affected Fluid Packs and discard them per local guidlines; and complete and return the enclosed faxback form and fax it to 1-888-943-5171 to receive replacement product for Fluid Packs from the affected lot. For questions contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-526-2272.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
CHL
Quantity affected
82 fluid packs
Distribution
US Nationwide Distribution in the states of LA, SC, CA, TX, NE, VA and WV including: Puerto Rico.
Date initiated
2015-07-14
Date posted
2015-08-21
Date terminated
2016-01-19
Location
Indianapolis, IN