COBAS AmpliPrep / COBAS TaqMan HCV Test, CE-IVD The COBAS AmpliPrep/COBAS TaqMan HCV Test is an in vitro nucleic acid am
FDA device recall Z-2587-2011 · posted 2011-06-17 · Terminated
Recall details
- Recalling firm
- Roche Molecular Systems, Inc.
- Reason for recall
- The COBAS Ampliprep/COBAS TaqMan HCV Test has been shown to under-quantitate a subset of genotype 4 patient specimens by approximately 1.0-1.5 log 10 in the absence of any sequence mismatches.
- Action taken
- The firm, Roche, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 14, 2011 and faxback form to their customers UPS ground. The letter described the product, problem and actions to be taken. The customers were instructed to review prior test results if there have been indications of discrepancies between test results generated with COBAS AmpliPrep/COBAS TaqMan HCV Test as compared to other relevant clinical and laboratory findings; complete and return the attached Product Advisory Notice: Affiliate Feedback Fax-Form via fax to: 1-866-792-5445, and file this letter for future reference, in particular for use with their 2010 CAP Survey Results to explain any under-quantitation of sample HVL 2-01. Please contact Roche Molecular Diagnostics Technical Support at 1-800-526-1247 if you have any questions about the information contained in this letter.
- Root cause
- Other
- Status
- Terminated
- Product code
- MZP
- Quantity affected
- 288,399 units
- Distribution
- USA (nationwide) and countries of: Australia, Austria, Azerbaijan, Belgium, Brazil, Canada, China, Croatia, Czech Republic, Egypt, Finland, France, Germany, Ghana, Greece, Hong Kong, India, Israel, It
- Date initiated
- 2011-04-14
- Date posted
- 2011-06-17
- Date terminated
- 2012-08-27
- Location
- Somerville, NJ