Cobas 8100 uni-directional reformatter (BRF) module with Software Version 02-xx

FDA device recall Z-1764-2017 · posted 2017-03-23 · Terminated

Recall details

Recalling firm
Roche Diagnostics Corporation
Reason for recall
Roche Diagnostics Corp. initiated a voluntary correction because a rack crash may occur on the Cobas 8100 bi-directional reformatter (BRF) and uni-directional reformatter (BRF) modules with software version 02-xx, only when rack buffering is activated and the rack buffer is completely full. This issue can lead to sample spillage, posing a potential risk to operators / laboratory staff due to exposure to potentially infectious material, cross-contamination of samples in the affected racks and erroneous results due to sample carry-over.
Action taken
Roche initiated a recall on 03/01/2017, notices were issued via UPS to customers and instructed customers to do the following: Roche is developing software version 03-01 to resolve this issue, and a Roche Representative will be contacting you in the near future to schedule an update. Until this software update is available, the rack buffering option for the BRF and URF modules must be disabled. If you are an authorized cobas 8100 automated workflow series operator, follow the steps outlined below to disable rack buffering. If your facility does not have an authorized cobas 8100 automated workflow series operator, please contact the Roche Support Network Customer Support Center at 1-800-428-2336 for further steps. If you have a major spill, set the affected module to offline and contact the Roche Support Network Customer Support Center at 1-800-428-2336. " If you are an authorized cobas 8100 automated workflow series operator, follow the steps outlined in this Urgent Medical Device Correction (UMDC) to disable rack buffering. " If your facility does not have an authorized cobas 8100 automated workflow series operator, please contact the Roche Support Network Customer Support Center at 1-800-428-2336 for further steps. " Complete the attached faxback form and fax it to 1-888-345-0024. " File this Urgent Medical Device Correction (UMDC) for future reference. Please contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-428-2336 if you have questions about the information contained in this UMDC.
Root cause
Under Investigation by firm
Status
Terminated
Product code
JQP
Quantity affected
45
Distribution
Domestic: MA, IN, AL, OH, MO, NE, LA, CA, IA, NJ, AR, PA, SC, IL, TX, and MI. Foreign: None
Date initiated
2017-03-01
Date posted
2017-03-23
Date terminated
2018-07-03
Location
Indianapolis, IN