CNSystems TASK FORCE MONITOR 3040i, plethysmograph, impedance. Noninvasively measure and display patient's hemodynamic p

FDA device recall Z-0572-2016 · posted 2016-01-04 · Terminated

Recall details

Recalling firm
Partners In Medicine Llc
Reason for recall
Device non-conformity which may lead to an incorrect oscillometric measurement of upper arm blood pressure values in some subjects - mainly in tall and slim persons (overestimation of systolic blood pressure, underestimation of diastolic blood pressure).
Action taken
CNSystems sent an Urgent Medical Device Recall Safety Notice dated December 1, 2015, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. CNSystems decided to perform a voluntary recall of the affected devices and to return the devices after repair (free of charge). For questions regarding this recall call 314-324-2470.
Root cause
Mixed-up of materials/components
Status
Terminated
Product code
DSB
Quantity affected
28 worldwide 2 monitors distributed in the U.S.
Distribution
Nationwide Distribution including MI and CA.
Date initiated
2015-12-01
Date posted
2016-01-04
Date terminated
2017-02-17
Location
Saint Louis, MO