Clinician Workstation Software, product code 2M5042
FDA device recall Z-0981-05 · posted 2005-07-13 · Terminated
Recall details
- Recalling firm
- Baxter Healthcare Corp.
- Reason for recall
- The medication administration results displayed in the 'View Results' link may be inaccurate.
- Action taken
- Urgent Device Correction recall letters dated 3/10/05 were sent to the direct accounts on the same date via first class mail. The letter describes the inaccuracies in the 'View Results' program, and informed the accounts that inaccurate information could lead to incorrect clinical decision-making, which could impact a patient''s medication treatment protocol. The accounts were requested to disable the 'View Results' link until a software upgrade is available, and to inform all clinical point of care users that the 'View Results' has been disabled and that the Manual Adminstration Record is to be used as the primary source of information for reviewing medication adminstration results. The accounts were instructed to contact their account executive and/or the Client Support Services at 1-877-835-2727 for instructions on how to disable the 'View Results' feature.
- Root cause
- Other
- Status
- Terminated
- Product code
- LNX
- Quantity affected
- 5 units
- Distribution
- Missouri, Massachusetts, Vermont, California and Canada.
- Date initiated
- 2005-03-10
- Date posted
- 2005-07-13
- Date terminated
- 2008-02-06
- Location
- Round Lake, IL