Clinical Chemistry Multiconstituent Calibrator--- List number: lE65-02

FDA device recall Z-0516-06 · posted 2006-02-14 · Terminated

Recall details

Recalling firm
Abbott Laboratories Diagnostic Div
Reason for recall
The two Multiconstituent Calibrator (MCC) lot numbers 19906M200 and 20879M100 failed to meet product performance expectations throughout the labeled expiration period, as a result of the shortened expiration dating, two lots are now expired. The expiration date for Lot number 26674M100 will be revised from 2006-07-31 to 2006-01-31.
Action taken
A Device Correction letter identifying the changes being made to the expiration dating for lot numbers 19906M200, 20879MI00 and 26674M1O0 along with a Customer Reply form will be distributed via FedEx to all AEROSET and ARCHITECT c8000 customers which utilize the Clincial Chemistry Multiconstituent Calibrator on December 27,2005. Outside the U.S. each country organization will need to determine the appropriate method for the effectiveness check.
Root cause
Other
Status
Terminated
Product code
JIX
Quantity affected
13961
Distribution
Nationwide plus worldwide in Mexico,Venezuela, Chile, Brazil, San Salvador, Singapore, Korea, Trinidad, Grand Cayman, Guatemala, Columbia, Uruguay, Puerto Rico, Canada, Hong Kong, Australia, Bermuda,
Date initiated
2005-12-23
Date posted
2006-02-14
Date terminated
2012-04-24
Location
South Pasadena, CA