Clinac Linear Accelerator; Model numbers H14, H27 and H29. Product Usage: Varian High Energy Linear Accelerator is indic
FDA device recall Z-2764-2011 · posted 2011-07-08 · Terminated
Recall details
- Recalling firm
- Varian Medical Systems, Inc. Oncology Systems
- Reason for recall
- The Coolant System input water supply manifold may leak in some Clinac Linear Accelerators, posing a risk of electrical shock to any person working within the protective housing.
- Action taken
- Varian Medical Systems, Inc. Oncology Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letters on May 18, 2011, via certified mail, return receipt requested. The letter identified the product, problem and recommended actions to be taken. The letter states that customers will be contacted by Varian Service to arrange for replacement of the recalled product. For additional information contact Varian Medical Systems, Inc at 888- 827-4265.
- Root cause
- Other
- Status
- Terminated
- Product code
- IYE
- Quantity affected
- 93
- Distribution
- Worldwide Distribution -- USA (nationwide) and countries of: Australia, Columbia, Denmark, France, Ireland, Italy, Japan, Morocco, Poland, Portugal, Saudi Arabia, South Korea, Spain, and United Kingdo
- Date initiated
- 2011-05-18
- Date posted
- 2011-07-08
- Date terminated
- 2012-03-13
- Location
- Palo Alto, CA