Clearlink System Continu-Flo Solution Set, 10 drops/mL, 113'', 2 Luer Activated Valves, Male Luer Lock Adapter, catalog

FDA device recall Z-1156-03 · posted 2003-08-21 · Terminated

Recall details

Recalling firm
Baxter Healthcare Corp.
Reason for recall
There are possible pinholes or perforations in the set tubing, caused by damage that occurred during the packaging process.
Action taken
Recalled by letter dated 7/16/03. The accounts were informed of the potential for leaks in the set tubing, and asked to inspect their inventory for the affected product codes and lots, segregate any affected products found and discard them according to local procedures. The customers will receive credit for the discarded affected product.
Root cause
Other
Status
Terminated
Product code
FPA
Quantity affected
887,483 units
Distribution
Nationwide, Mexico, Canada, Columbia, Australia, New Zealand and Taiwan
Date initiated
2003-07-16
Date posted
2003-08-21
Date terminated
2004-06-29
Location
Round Lake, IL