Cleaning probe included as an accessory to aspirators that are sterile packed single use disposable accessories to the G

FDA device recall Z-1130-2007 · posted 2007-08-07 · Terminated

Recall details

Recalling firm
GE OEC Medical Systems, Inc
Reason for recall
Cleaning probes used to unclog disposable aspirators may shed small particles of nylon material which may be transferred into the nasal cavity.
Action taken
Consignees were notified by certified mail on 05/18/2007 and were told to stop using and to dispose of the affected probes.
Root cause
Other
Status
Terminated
Product code
LLZ
Quantity affected
67,625 units
Distribution
Worldwide, including USA, Canada and France.
Date initiated
2007-05-18
Date posted
2007-08-07
Date terminated
2011-08-22
Location
Salt Lake City, UT