CIVCO Surgi-Tip Transducer Cover, Ref. 610-834, Sterile (20.3 x 244cm) polyethylene telescopically-folded cover with att

FDA device recall Z-2242-2010 · posted 2010-08-20 · Terminated

Recall details

Recalling firm
Civco Medical Instruments Inc
Reason for recall
Polyethylene material did not have proper treatment applied during manufacturing. Untreated polyethylene material does not maintain a secure seal with the adhesive tape used to secure the tip to the cover.
Action taken
CIVCO sent a "Urgent - Medical Device Correctdion" letter date July 19, 2010 to consignees describing the reason for recall and requesting they dispose of the affect lot number of product. A response form was enclosed for completion and return to the firm, indicating the amount of product in inventory and requesting a Return Authorization Number. The form was to be returned via FAX or e-mail. Customers may contact CIVO at 800-445-6741.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
ITX
Quantity affected
22 boxes
Distribution
Nationwide distribution: the states of OH, MD, NE, VA, IA, MI, WA, and WI.
Date initiated
2010-07-20
Date posted
2010-08-20
Date terminated
2012-04-13
Location
Kalona, IA