Atlas FDA

CIVCO SBRT System, REF MTSBRT001L, containing CIVCO Pro-Lok Platform, REF MTSBRT002. Firm name on label is CIVCO, Orange

FDA device recall Z-3021-2011 · posted 2011-08-15 · Terminated

Recall details

Recalling firm
Civco Medical Instruments Inc
Reason for recall
Systems sold before December 2009 displayed a scale of 0-40cm going from left to right, while systems after December 2009 reversed this and were displayed as 40-0cm from left to right.
Action taken
CIVCO issued a "FIELD SAFETY NOTIFICATION" letter dated 12/20/2010 beginning 12/30/2010, via FedEx or UPS to distributors and customers. The notification identifies the affected device, problem, and field safety corrective action. The letter states that if customers have more than one system purchased from different time periods that they should examine their systems and assess if there is any mismatch. The recalling firm will rework the system to correct the difference and send it back to the customer. To take advantage of this correction customers are advised to contact the recalling firm within thirty days of receiving the safety notification. The letter includes a response form to be completed and returned as per instructions, to indicate that the notice has been read and understood by all operators of the equipment. The letter to distributors requested that a copy of the notification letter be forward to all customers who have received this component. For assistance contact your CIVCO insides sales representative by calling CIVCO Medical Solutions, Radiation Oncology, at 800-842-8688.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
IXQ
Quantity affected
49 systems
Distribution
Worldwide Distribution - USA ( nationwide) including the state of PA. and countries of: Canada, China, England, Israel, Italy, Japan, Korea, The Netherlands, Russia, Spain, Sweden, Switzerland and Tur
Date initiated
2010-12-30
Date posted
2011-08-15
Date terminated
2013-06-28
Location
Kalona, IA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
CARBON FIBER CONFORMAL COUCH TOP (K973842)Medtec, Inc.1997-12-05