Civco, MTSR02 Shoulder Retractor with Sherpa Wrist Loops Kit with Base, Non-sterile re-usable wool wrist strap, REF: 20S
FDA device recall Z-0989-2015 · posted 2015-01-15 · Terminated
Recall details
- Recalling firm
- Med Tec Inc
- Reason for recall
- The product (wool wrist loop) was manufactured 7cm shorter than product specifications.
- Action taken
- The firm sent e-mailed Field Safety Notice letters and contacted customers via phone on December 11, 2014, informing customers of the issue with the product. Customers were advised they may accept the product as is or request a replacement. A response form was enclosed for consignees to indicate the status of the product. Please contact a CIVCO insides sales representative for additional assistance. You may reach your representative by calling CIVCO Medical Solutions, Radiation Oncology, at 800-842-8688 or 712-737- 8688 between 8:00AM and 5:00PM CST.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- FMQ
- Quantity affected
- 11
- Distribution
- Worldwide Distribution - US Distribution to the states of : AZ, MO, PA and WI., and the countries of : Austria, India and Japan.
- Date initiated
- 2014-12-11
- Date posted
- 2015-01-15
- Date terminated
- 2017-02-03
- Location
- Orange City, IA