Civco Belly Board, Reusable non-sterile carbon fiber belly board with cushion, REF 125012, packaged 1 per box. Product U

FDA device recall Z-0802-2014 · posted 2014-01-21 · Terminated

Recall details

Recalling firm
Civco Medical Instruments Inc
Reason for recall
Scales adhered to the product may be misaligned from the zero reference such that one side will not match the other.
Action taken
Civco Medical Instruments Inc sent a FIELD SAFETY NOTIFICATION letter dated October 28, 2013 via UPS to their customers. The notification identified the affected device, problem, and field safety corrective action. The letter describes how customers can check to see whether the scales on their device are out of alignment. The letter states CIVCO will rework the system to correct the difference. For questions contact sales representative at CIVCO Medical Solutions, Radiation Oncology at 800-842-8688 or 712-737-8688.
Root cause
Device Design
Status
Terminated
Product code
IYE
Quantity affected
138 belly boards
Distribution
Worldwide Distribution - USA Nationwide in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, MA, MD, MI, MN, MT, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, and WV. Foreign distribution was ma
Date initiated
2013-10-25
Date posted
2014-01-21
Date terminated
2017-02-02
Location
Kalona, IA