Circular Stapler 21 mm Digital Loading Unit, product code CS21; Circular Stapler 25 mm Digital Loading Unit, product cod

FDA device recall Z-0393-2010 · posted 2009-11-23 · Terminated

Recall details

Recalling firm
Surgical Devices/Covidien
Reason for recall
Lack of assurance that products were manufactured under GMPs. This action is being undertaken due to certain products potentially exhibiting compromised staple formation or cutting which in some instances may result in extended surgical time for the patient.
Action taken
The recalling firm issued an Urgent Medical Device Recall letter dated 9/22/09 to all customers. The letter informed customers that Covidien acquired PMI, about the problem with the products, and about the need to return any products remaining inventory. Customers were asked to return the entire inventory of Power Medical Interventions product, regardless of its expiration date. This process can be initiated by contacting Power Medical Innovations customer service at 1-866-769-3763.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
GDW
Quantity affected
38,058 units
Distribution
The products were shipped medical facilities nationwide. The product was also shipped to distributors in Canada, Japan, Germany, Ireland, Italy, Spain, Taiwan, Belgium, Sweden, Netherlands, Israel, Po
Date initiated
2009-09-23
Date posted
2009-11-23
Date terminated
2010-03-08
Location
North Haven, CT