CircuCool Fluid Circulation Pump used in the Chilli Cooled Ablation System, Model Number M00480050, manufactured by Barn

FDA device recall Z-1156-2010 · posted 2010-03-17 · Terminated

Recall details

Recalling firm
Boston Scientific Corp
Reason for recall
Product failure-- Product may cease operation as a result of potential speed control board failure when the device is shut off, and may not turn back on. A component on the circuit board fails unexpectedly because of a power surge, causing a diode to burn out.
Action taken
The firm has issued notification letters via Federal express with reply verification tracking forms. A follow up letter and telephone survey as needed will be sent to follow up with non-responders. The firm is asking its consignees to return any recalled product. If there are any questions, please call 1.408.935.4952.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
LPB
Quantity affected
75 units
Distribution
US and to Spain, France, Germany, Italy, Russia, Poland, the Netherlands, Israel and Belgium
Date initiated
2009-09-22
Date posted
2010-03-17
Date terminated
2010-12-27
Location
San Jose, CA