Circlip component of the Stryker Flat Panel and Navigation Arm System. Stryker Flat Panel and Navigation Arm System are
FDA device recall Z-0182-2009 · posted 2008-10-22 · Terminated
Recall details
- Recalling firm
- Stryker Communications Corp
- Reason for recall
- Circlip component used to suspend flat panel and navigation arm system may become dislodged.
- Action taken
- Notification letter was sent to consignees September 5, 2008. Field service will be performed at all affected sites. Old Circlips will be replaced with new clamps that do not allow the arm to fall. Contact Stryker Communications Corp. at 1-972-410-7310 for assistance.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- FQO
- Quantity affected
- 705 units
- Distribution
- Nationwide and International Distribution
- Date initiated
- 2008-09-05
- Date posted
- 2008-10-22
- Date terminated
- 2009-10-02
- Location
- Flower Mound, TX