Cholestech LDX Lipid Profile"GLU test cassettes, Catalog Number 10-991, Lot # 245104 Cholestech LDX Lipid-Glucose tests

FDA device recall Z-2402-2011 · posted 2011-06-01 · Terminated

Recall details

Recalling firm
Biosite Inc Dba Innovacon Inc.
Reason for recall
The recall was initiated due to an issue related to an error in the encoding of calibration coefficients in the sample cassette that cause the Cholestech LDX analyzer to report glucose values as >500 mg/dL regardless of the true value of the specimen.
Action taken
Alere San Diego, Inc ( formerly known as Biosite Incorporated sent an URGENT MEDICAL DEVICE RECALL letter on February 18, 2011, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. Customers were requested to: (1) Discontinue use and/or sale of Cholestech LDX Lipid Profile-GLU test cassettes, Catalog Number 10-991, Lot # 245104. (2) Discard in accordance with your local regulations, all units of the above listed product. (3) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. (4) If you have forwarded the product listed above to another laboratory, please provide a copy of this letter to them. (5) Complete and FAX the enclosed Verification Form within 10 days to confirm your receipt of this notice, to indicate the number of boxes of test cassettes discarded from your inventory and replacement product will be sent to you. For questions about the information contained in this notification, customers were instructed to contact Alere Technical Services at 1-877-308-8289 (24 hours a day, 7 days a week) or via e-mail at service.cholestech@alere.com.
Root cause
Pending
Status
Terminated
Product code
CGA
Quantity affected
5,305 Boxes
Distribution
Worldwide Distribution - USA ( nationwide ) including the countries of South America ( EL Salvador) , Europe ( Denmark, France, Norway and Switzerland), Bahrain, Kuwait, South Africa, and United Arab
Date initiated
2011-02-18
Date posted
2011-06-01
Date terminated
2012-04-02