Atlas FDA

CERCLAGE WIRE, 0.6 MM DIA. X 175 MM LENGTH, SST, REF 71163006, QTY: (6), STERILE R, CE 008, 2020/07, Smith & Nephew, Inc

FDA device recall Z-2197-2010 · posted 2010-08-10 · Terminated

Recall details

Recalling firm
Smith & Nephew Inc
Reason for recall
Plastic trays which encase these products have burrs on the hinged packaging which can puncture the single barrier sterile pouch, compromising sterility.
Action taken
All affected Smith & Nephew Sales Representatives were notified of problem and the recall was sent via overnight mail on 01/23/2009. All affected Smith & Nephew International Distributors were notified via email and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return.
Root cause
Process control
Status
Terminated
Product code
JDQ
Quantity affected
1 unit
Distribution
KY, TX, DC, NC, CA, AZ, GA and OH; 3 International: Australia, Malaysia, and Switzerland
Date initiated
2010-06-28
Date posted
2010-08-10
Date terminated
2012-01-23
Location
Memphis, TN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.