Centrifuge Model CF-6 with lid lock. (Block Model number OCT-VI). Units are distributed to Bloch Scientific, Holbrook, N
FDA device recall Z-0563-06 · posted 2006-02-28 · Terminated
Recall details
- Recalling firm
- Cygnus LLC
- Reason for recall
- The units were exhibiting tolerance stacking in the electronic circuit timer that controls the electric motor that spins the centrifuge and applies the centrifuge brake.
- Action taken
- A recall notificaiton was sent to the firm''s customers by e-mail on 2/14/2005. Distributor conducted sub-recall.
- Root cause
- Other
- Status
- Terminated
- Product code
- JQC
- Quantity affected
- 20 units
- Distribution
- All product was shipped domestically to the firm''s distributor, Bock Scientific, 1170 Lincoln Avenue, Unit 11, Holbrook, NY 11741
- Date initiated
- 2005-02-04
- Date posted
- 2006-02-28
- Date terminated
- 2007-02-01
- Location
- Paterson, NJ