Centricity Universal Viewer Product Usage: The Centricity Universal Viewer Version 6.0 software is a picture archiving a
FDA device recall Z-2470-2015 · posted 2015-08-27 · Terminated
Recall details
- Recalling firm
- GE Healthcare
- Reason for recall
- Images from the Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer may be missing when a system parameter MapRoute is set to a value greater than 1.
- Action taken
- The firm sent out an Urgent Medical Device Correction letter dated 8/11/2015 to all affected customers. Letter states that No User activity required. GE Healthcare has completed a remote inspection of your system and found that this system parameter is currently set to the default setting of 1. The firm will correct all affected systems by a providing software update. The letter also states that if the customer has any questions they should call GE Healthcare Service at 1-800-437-1171 or their local service representative.
- Root cause
- Software design (manufacturing process)
- Status
- Terminated
- Product code
- LLZ
- Quantity affected
- 26 systems
- Distribution
- Worldwide Distribution - U.S. nationwide in the states of: AK, CA, CO, DE, FL, HI, ID, NY, PA, SD, TN and TX; France, Germany, Greece, India, Israel, Malaysia, Mexico, Northern Ireland (UK), Panama, P
- Date initiated
- 2015-08-11
- Date posted
- 2015-08-27
- Date terminated
- 2017-01-05
- Location
- Barrington, IL