Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from o

FDA device recall Z-0819-2020 · posted 2020-01-17 · Terminated

Recall details

Recalling firm
GE Healthcare, LLC
Reason for recall
Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
Action taken
The firm initiated their correction on 12/16/2019 by letter. The letter explained the issue and directed the consignee to follow instructions in the letter until an update to the Centricity Universal Viewer can be made by the firm to correct the issue.
Root cause
Software design
Status
Terminated
Product code
LLZ
Quantity affected
205
Distribution
Distributed nationwide; Internationally distributed to Mexico and 24 additional countries.
Date initiated
2019-12-16
Date posted
2020-01-17
Date terminated
2023-10-05
Location
Waukesha, WI