Centricity PACS System versions 3.X and higher and 4.X and higher Product Usage: The Centricity PACS Workstation is inte

FDA device recall Z-0731-2014 · posted 2014-01-14 · Terminated

Recall details

Recalling firm
Ge Healthcare It
Reason for recall
There is a potential loss of image(s) associated with the Centricity to Centricity (C2C) exam transfer module of Centricity PACS System versions 3.X and higher and 4.X and higher when transferring from the source to a destination server. When another process in the destination server attempts to access the same object or table, the transfer process of a particular image may be terminated. Once ter
Action taken
GE Healthcare sent an Important Product Information letter dated September 15, 2012 to affected customers. The letter identified the affected product problem and actions to be taken. GE Healthcare will provide a patch to 3.2.X and 4.0 Systems to remediate the issue. For questions contact your local GEHC-IT service representative.
Root cause
Software design
Status
Terminated
Product code
LLZ
Quantity affected
748 units
Distribution
Worldwide Distribution - US: Nationwide (AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX,
Date initiated
2012-09-15
Date posted
2014-01-14
Date terminated
2016-06-10
Location
Barrington, IL