CELL-DYN Emerald Cleaner, product number 09H46-02. 960mL. IVD; Hematology: The CELL-DYN Emerald Cleaner is designed for

FDA device recall Z-1522-2016 · posted 2016-04-19 · Terminated

Recall details

Recalling firm
Abbott Laboratories
Reason for recall
Three lots of CELL-DYN Emerald Cleaner may have occurrences where the analyzer generates Quality Control Out of Range Low for parameters RBC and PLT.
Action taken
All customers who were shipped the affected lots were sent a Product Recall letter on March 24, 2016 via Federal Express. Letters were titled Product Recall - Immediate Action Required. An additional four lots were discovered to be affected by the recall and the firm send letters to customers affected by these lots on June 27, 2016
Root cause
Under Investigation by firm
Status
Terminated
Product code
GKZ
Quantity affected
10,364 plus 9410 additional units for recall expansion.
Distribution
Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Austria, Bosnia & Herze, Cameroon, Egypt, Germany, Indonesia, Iraq, Lebanon, Mali, Montenegro, Netherlands, Philippine
Date initiated
2016-03-24
Date posted
2016-04-19
Date terminated
2017-01-04
Location
Santa Clara, CA