Atlas FDA

CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial K

FDA device recall Z-2018-2025 · posted 2025-06-24 · Open, Classified

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.
Action taken
On May 28, 2025, firm began notifying affected consignees via Urgent Medical Device Removal letters. A Terumo Service Representative will contact each customer to discuss removal and replacement of affected devices. The firm is not recommending that customers halt use of their CDI OneView BPM Probe while waiting for removal and replacement. In the event suspected parameter inaccuracy occurs, performance of an in vivo recalibration is recommended. Per the Operator's Manual, displayed values should be verified for accuracy from another source before initial treatment.
Root cause
Employee error
Status
Open, Classified
Product code
DRY
Quantity affected
28 US; 7 OUS
Distribution
US Nationwide distribution in the states of California, Delaware, Hawaii, Iowa, Indiana, Michigan, North Carolina, New York, Ohio, and Washington. The country of New Zealand.
Date initiated
2025-05-28
Date posted
2025-06-24
Location
Ann Arbor, MI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
CDI OneView Monitoring System (K234065)Terumo Cardiovascular Systems Corporation2024-04-25