CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial K
FDA device recall Z-2018-2025 · posted 2025-06-24 · Open, Classified
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.
- Action taken
- On May 28, 2025, firm began notifying affected consignees via Urgent Medical Device Removal letters. A Terumo Service Representative will contact each customer to discuss removal and replacement of affected devices. The firm is not recommending that customers halt use of their CDI OneView BPM Probe while waiting for removal and replacement. In the event suspected parameter inaccuracy occurs, performance of an in vivo recalibration is recommended. Per the Operator's Manual, displayed values should be verified for accuracy from another source before initial treatment.
- Root cause
- Employee error
- Status
- Open, Classified
- Product code
- DRY
- Quantity affected
- 28 US; 7 OUS
- Distribution
- US Nationwide distribution in the states of California, Delaware, Hawaii, Iowa, Indiana, Michigan, North Carolina, New York, Ohio, and Washington. The country of New Zealand.
- Date initiated
- 2025-05-28
- Date posted
- 2025-06-24
- Location
- Ann Arbor, MI