CCX Operating Software , Version 4.08
FDA device recall Z-0578-06 · posted 2006-02-28 · Terminated
Recall details
- Recalling firm
- Nova Biomedical Corporation
- Reason for recall
- Analyzers may fail to perform scheduled automatic two-point calibrations and Quality Control runs
- Action taken
- Nova Biomedical Issued on 1/27/2006 an Advisory Notice to Stat Profile CCX Customers to report the malfunction and request the Good Laboratory Practices (GLP) be used and review be made to assure the calibration status and QC is being performed. A new software version is scheduled to be released to correct the error.
- Root cause
- Other
- Status
- Terminated
- Product code
- CHL
- Quantity affected
- 33 units
- Distribution
- Nationwide Foreign: Canada, UAE, Taiwan, Singapore, Budapest, UK, France, Malaysia, France, Mexico, Korea, Finland, Germany, Turkey, Ireland, Italy, Austria, Australia, Netherlands, Venezuela, Japan,
- Date initiated
- 2006-01-27
- Date posted
- 2006-02-28
- Date terminated
- 2006-10-19
- Location
- Waltham, MA