Catalog # 9024-00 or LSW-S-3040 for 4.5mm BONE-LOK PLS Implant A sterile, canulated screw that is supplied as a stand-al
FDA device recall Z-3034-2011 · posted 2011-08-18 · Terminated
Recall details
- Recalling firm
- Interventional Spine Inc
- Reason for recall
- This recall has been initiated due to Pull Pins that may disengage from the Facet Screw during the compression step, requiring that compression be completed with a device driver rather than the compression tool. Use of this product may result in less than optimal compression leading to pseudoarthrosis, non-union or facet pain.
- Action taken
- Interventional Spine, Inc. sent a recall letter dated March 15, 2011, informing consignees of the product, the problem, and the action to be taken. Consignees were asked to immediately examine their inventory and quarantine product subject to recall. In addition, if they distributed the product further they were asked to notify Jane Metcalf, the Vice President of Quality Assurance at once. Interventional Spine further informed consignees that the recall was being carried out to the individual consignee level. And their assistance was appreciated and necessary to prevent use of this product. They were then asked to complete and return the enclosed response form as soon as possible. Questions were directed to call their regional directors. For questions regarding this recall call 949-472-0006.
- Root cause
- Other
- Status
- Terminated
- Product code
- MRW
- Quantity affected
- 407 units
- Distribution
- Nationwide Distribution including CO, CT, FL, KY, LA, MA, MO, NC, NY, OK, OR, TN, TX, VA & WI
- Date initiated
- 2011-03-15
- Date posted
- 2011-08-18
- Date terminated
- 2013-02-11
- Location
- Irvine, CA