CARTO 3 System Interface Cable, 34-pin to 10-pin, Red, 10' - Part Number: D-1286-18; Catalog Number CR3410CT This produc

FDA device recall Z-0884-2011 · posted 2011-01-07 · Terminated

Recall details

Recalling firm
Biosense Webster, Inc.
Reason for recall
When connecting this cable to the mapping system, loss of Body Surface ECG signals may occur. This may result in inaccurate locations appearing for catheters displayed 3D on the CARTO monitor.
Action taken
Biosense Webster issued an Urgent Field Safety Notice letter dated October 5, 2010. The letter identified the product, the problem,and the action to be taken by the customer. Customers were instructed to return the affected cable, accompanied by the attached Return Authorization form. For questions regarding this recall call Biosense Webster at 909 839-8874.
Root cause
Device Design
Status
Terminated
Product code
DSA
Quantity affected
1692
Distribution
Worldwide Distribution: USA, including the states of: AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, KS, KY, MA, MD, MI, MN, NC, NE, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA, WI, and WV, and the countr
Date initiated
2010-10-05
Date posted
2011-01-07
Date terminated
2012-05-07
Location
Irwindale, CA