Atlas FDA

Carroll Healthcare brand Arro bed models BA3180-151 AE, BA3180-151 PE, BA3180-151 FE, BA3180-171 AE, BA3180-171 FE and B

FDA device recall Z-0919-05 · posted 2005-06-22 · Terminated

Recall details

Recalling firm
Carroll Healthcare Inc.
Reason for recall
The diodes in the control panel may fail, which could result in the thermal breakdown of the circuit board inside of the control box, and cause the bed to cease functioning.
Action taken
Consignees were sent a fax on 5/17/05 advising them of the problem and instructing them to lock-out the HILO pendant function on the attendant control pad until the firm visits their facility and corrects their beds.
Root cause
Other
Status
Terminated
Product code
FNL
Distribution
Nationwide.
Date initiated
2005-05-17
Date posted
2005-06-22
Date terminated
2005-11-17
Location
London, Ontario, Canada