Cardiovascular Procedure Kit, P/N 64789-01, Wisconsin Cardiovascular Services Pump and Table Pack, Terumo Cardiovascular
FDA device recall Z-1311-2008 · posted 2008-03-28 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corp
- Reason for recall
- Loose Flash at tip of Weighted Flexible Sucker: A loose strand of stainless steel left on the device during manufacturing by an outside supplier may result in an unretrieved metal fragment remaining in the patient.
- Action taken
- Consignees were notified via recall letter, Urgent Medical Device Removal, dated 1/29/08 via fax or email to cease using the weighted sucker in the packs and to use an alternative sucker until replacement product arrives from Terumo.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- OEZ
- Quantity affected
- 110
- Distribution
- Nationwide-USA including states of Missouri, Tennessee and Wisconsin-Custom Kits for 3 hospitals.
- Date initiated
- 2008-01-29
- Date posted
- 2008-03-28
- Date terminated
- 2009-07-07
- Location
- Ann Arbor, MI