Cardiovascular Procedure Kit containing Disposable Centrifugal Pumphead with or without X-Coating, sterile single use. C
FDA device recall Z-2030-2014 · posted 2014-07-11 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.
- Action taken
- Urgent Medical Device Recall letters, dated April 11, 2014, were sent to customers. The letters identified the affected product, as well as the reason for the correction, the potential hazard, the correction, the affected population, and instructions for customers. Customers are to review the notice and ensure that all users have received a copy. Additionally, customers are to confirm receipt of the communication by returning the attached Customer Response Form as instructed; and, return all effected centrifugal pumps. Terumo CVS will issue a Returned Goods Authorization upon receipt of the Response Form. Questions should be directed to Terumo CVS Customer Service at 800-521-2818.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- KFM
- Quantity affected
- 256 kits
- Distribution
- Nationwide Distribution
- Date initiated
- 2014-04-14
- Date posted
- 2014-07-11
- Date terminated
- 2016-09-07
- Location
- Elkton, MD
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| SARN CENTRIFUGAL PUMP (K112229) | Terumo Cardiovascular Systems | 2011-12-09 |