CAPNOCHECK BCI Airway Adapter; Model Number BSR; Product Code WW1100

FDA device recall Z-1404-2025 · posted 2025-03-20 · Open, Classified

Recall details

Recalling firm
Smiths Medical ASD, Inc.
Reason for recall
Airway adapter does not meet specification and may break off when attaching a valve or tube. Device failure may result in inaccurate readings, a break in airway circuit, and loss of ventilation. A break in the airway circuit could result in inability to ventilate, hypoxia, cardiopulmonary collapse, or death.
Action taken
On February 14, 2025, customers were notified via Urgent Medical Device Correction letters. Consignees were instructed to locate and quarantine all affected product within their institution and destroy them. The recalling firm will contact customers who have returned completed response forms identifying affected products within their possession to discuss credit.
Root cause
Process control
Status
Open, Classified
Product code
CCK
Quantity affected
175,994 US; 45,223 OUS
Distribution
US Nationwide distribution. International distribution pending.
Date initiated
2025-02-14
Date posted
2025-03-20
Location
Minneapolis, MN