Atlas FDA

Capillary (end) caps, accessory for use with blood gas analyzers included in Radiometer''s CLINITUBES kits. The end caps

FDA device recall Z-1322-06 · posted 2006-08-03 · Terminated

Recall details

Recalling firm
Radiometer America Inc
Reason for recall
Capilliary end caps, an accessory for ABL blood gas analyzers, leak blood from capillary tubes during transport.
Action taken
The firm notified their customers of this recall action with a letter dated 3/22/2006. The letter was accompanied by a single bag of new end caps. The letter informs the customers of the ''leakage'' problem and asks that the customer examine their inventories and discard any of the referenced end caps found on-hand at any of their various locations. The letter also includes a response form to be faxed back to Radiometer indicating the amount of replacement capillary end caps required by the customer in addition to the one bag received. The recalling firm also notified the customers that the suspect end caps will all be replaced ASAP.
Root cause
Other
Status
Terminated
Product code
GIO
Quantity affected
Green Caps: 208 kits. Each kit contains 1 to 5 bags of caps.
Distribution
Nationwide, Canada and Istanbul.
Date initiated
2006-03-22
Date posted
2006-08-03
Date terminated
2012-02-24
Location
Westlake, OH