Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C.

FDA device recall Z-2265-2012 · posted 2012-08-24 · Terminated

Recall details

Recalling firm
Canon Inc.
Reason for recall
Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.
Action taken
Canon Inc. notified sales companies including Canon USA on 9/15/2010 with Service Bulletin 10-017, and on 8/25/2011 with Service Bulletin 11-011, and on 1/20/2011 with Service Bulletin 11-001. Sales representatives were to visit customers and check and replace the screws.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
MQB
Quantity affected
468 units
Distribution
Nationwide Distribution (USA)
Date initiated
2010-11-01
Date posted
2012-08-24
Date terminated
2014-01-07
Location
Ohta-ku, Tokyo, Japan