Cambridge Endoscopic Autonomy Laparo-Angle 5mm Metzenbaum Scissors with Monopolar Cautery, 45 cm Catalog Number: LA1101-

FDA device recall Z-2744-2011 · posted 2011-07-06 · Terminated

Recall details

Recalling firm
Cambridge Endoscopic Devices, Incorporated
Reason for recall
PET shrink sleeve insulation material dislodging from the distal tips of Autonomy Laparo-Angle (ALA) Metzenbaum Scissors
Action taken
Cambridge Endoscopic Devices, Inc. sent an "IMPORTANT MEDICAL DEVICE RECALL" letter dated May 20, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. Customers are instructed to return unused inventory using RGA Number 301. Returned devices will be replaced. Questions regarding this recall, call 508-405-0790 for assistance.
Root cause
Process control
Status
Terminated
Product code
GEI
Quantity affected
255 units
Distribution
Worldwide Distribution--USA (nationwide) and the countries of Armenia , Australia, Austria, Chile, China, Denmark, Hong Kong, India, Ireland , Italy, Japan, Saudi Arabia, South Korea, Spain, Switzerla
Date initiated
2011-05-20
Date posted
2011-07-06
Date terminated
2012-07-24
Location
Framingham, MA