FDA device recall Z-1937-2010 · posted 2010-06-30 · Terminated

Recall details

Recalling firm
Centurion Medical Products
Reason for recall
The 3M micropore tape is under recall due to the tape not performing and needles becoming loose and dislodged.
Action taken
The firm sent Urgent Recall Product Notices to its two customers dated 3/16/2010. The letter explained that the tape may not adhere properly. The consignees were instructed to remove any kits containing the recalled tape from their inventory, if the tape was used to secure dialysis needles or pressure dressings. If the tape was used for other purposes, the kits could still be used. The recall letter included a reply form for customers to complete and email or fax back. Upon receipt of the form, a Centurion Representative will contact them to retrieve affected product. Customers should contact Matthew Price at 517-546-5400 ext. 1135, if further info is needed.
Root cause
Other
Status
Terminated
Product code
FKG
Quantity affected
24 kits
Distribution
Nationwide Distribution -- AR and OH.
Date initiated
2010-03-16
Date posted
2010-06-30
Date terminated
2011-02-09
Location
Howell, MI