Bubble Sensor for 3/8" x 3/32" tubing; Model Number: 701055720;

FDA device recall Z-1300-2026 · posted 2026-02-05 · Open, Classified

Recall details

Recalling firm
Maquet Cardiopulmonary Gmbh
Reason for recall
Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Venous Bubble Sensor. Excessive bending of the connecting cable can lead to full damage of the Venous Bubble Sensor or loose cable contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device. These errors can occur temporarily when the connecting cable is moved or permanently if the connection is fully compromised.
Action taken
On or around January 9, 2026, "URGENT MEDICAL DEVICE RECALL" letters were sent to customers. Customers were instructed to take the following actions: - Please examine your inventory immediately to determine if you have any affected product in your inventory and utilize the recommended mitigations listed above until replacement sensors are available. - Complete and return the enclosed response form Getinge by e-mailing a copy to recallresponses.qrc@getinge.com. - Contact Customer Service via email at ACTSalesSupport.US@Getinge.com or 888-943-8872, ext 28849 / (888 9GETUSA), Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. (Eastern Time Zone) to order a replacement sensor that will be shipped once available. Affected sensors will be returned to Getinge at time of replacement. - Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. If you have any questions, contact Customer Service via email at ACTSalesSupport.US@Getinge.com or 888-943-8872, ext 28849 / (888 9GETUSA), Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. (Eastern Time Zone).
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
DTQ
Quantity affected
3050 units
Distribution
Worldwide Distribution: US (Nationwide) and OUS (International) to countries of: Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Bosnia and Herzegovina, Botswana, Brazil
Date initiated
2026-01-09
Date posted
2026-02-05
Location
Rastatt, Germany