BS-200 Chemistry Analyzer Designed for clinical laboratory use.

FDA device recall Z-0977-2013 · posted 2013-03-22 · Terminated

Recall details

Recalling firm
Mindray DS USA, Inc. d.b.a. Mindray North America
Reason for recall
An issue has been identified with the BS-200 where clinical chemistry results may be inaccurate.
Action taken
Mindray North America /Mindray DS USA Inc. sent Urgent Product Corrective Action letters via certified mail, return receipt requested, signature required on 11/28/2012. The letters identified the affected product as well as the issue, effect on patients, and the corrective action. Customers are to remove the analyzer from service until the affected pump has been replaced. Also, customers are to contact Mindray Customer Service Representative Tony Li at 1-800-288-2121 x2112 for technical assistance, and to arrange for replacement pumps at no charge.
Root cause
Device Design
Status
Terminated
Product code
CFR
Quantity affected
3 units
Distribution
Distributed in the states of NC and TX.
Date initiated
2012-11-28
Date posted
2013-03-22
Date terminated
2013-06-12
Location
Mahwah, NJ