Atlas FDA

Broselow Pediatric Emergency Rainbow Tape (distribution by AirLife); REF 7700REA AirLife brand, 2025 Edition, and 36-234

FDA device recall Z-1904-2025 · posted 2025-06-20 · Open, Classified

Recall details

Recalling firm
SunMed Holdings, LLC
Reason for recall
The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of joules, causing significant harm including burns, heart damage, and cardiac arrest. Incorrect values in the Orange zone could lead to overdosing the patient with sodium bicarbonate and cause metabolic alkalosis, electrolyte imbalances, tissue damage, and worsen respiratory status. Incorrect information in the Grey may lead to underdosing the patient with sodium bicarbonate and may cause reduced myocardial contractility, decreased response to vasopressors, and increased risk of dysrhythmia.
Action taken
On May 15, 2025, the firm began notifying customers via Urgent Medical Device Recall letters. Customers were given detailed information about the incorrect information printed on the tape. Customers were instructed to identify and quarantine all affected product in inventory. Customers should destroy and dispose of all affected product. If product was further distributed, distributors should identify their customers/consignees and notify them of the recall. The firm will provide replacement products once customer confirms destruction of affected product.
Root cause
Process control
Status
Open, Classified
Product code
OKI
Quantity affected
9900 eaches
Distribution
US distribution nationwide. International distribution to Canada.
Date initiated
2025-05-15
Date posted
2025-06-20
Location
Grand Rapids, MI

Monitor recalls programmatically