Brilliance Big Bore - Model #728243, and - Model #728244, CT Scanning System is subject to recall/correction Product Usa

FDA device recall Z-1692-2012 · posted 2012-05-30 · Terminated

Recall details

Recalling firm
Philips Medical Systems (Cleveland) Inc
Reason for recall
Component defect. Following their receipt of a customer complaint, Philips Medical determined that the metal blocks which support the scissor slide rails underneath the scanning table (couch) of several of the firm's CT scanning systems were not manufactured to specification and may crack.
Action taken
Philips issued an 'URGENT Medical Device Correction' letter dated March 27, 2012 to all affected customers. The letter identified the affected products, problem, hazard involved, how to identify affected products, and actions to be taken by customer/user. The letter informed customers that Philips intends to dispatching service personnel to each customer site in order to correct this issue. In the meantime, the letter recommend customer to contact their Philips representative if they hear unusual noises when moving the couch vertically, or if there is unintended vertical motion of the couch. For questions contact your local Philips representative or local Philips Healthcare office. For North America and Canada contact the Customer Care Solutions Center 1-800-722-9377, option 5: Enter Site ID or follow the prompts.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JAK
Quantity affected
12 system units were distributed in the U.S.
Distribution
Worldwide Distribution - U.S. (Nationwide) including the states of: CA, FL,, HI, IA, IN, KY, LA, MA,, MI, MO, MS. NM, NV, OH, PA, PA, SC, TN, TX, WA, and Puerto Rico
Date initiated
2012-03-26
Date posted
2012-05-30
Date terminated
2013-02-22
Location
Cleveland, OH

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
BRILLIANCE VOLUME (K060937)Philips Medical Systems (Cleveland), Inc.2006-06-05
ACQSIM-MULTISLICE-CT, MODEL 453567042271 (K033357)Philips Medical Systems (Cleveland), Inc.2004-01-09
PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM (K033326)Philips Medical Systems, Inc.2003-10-29
MX8000 IDT CT SYSTEM (K012009)Marconi Medical Systems, Inc.2001-09-13