Atlas FDA

Brilliance 64, Computed Tomography X-Ray System, Model Numbers: 728231 and 728326, including the Ingenuity CT. The recal

FDA device recall Z-2206-2011 · posted 2011-05-19 · Terminated

Recall details

Recalling firm
Philips Medical Systems (Cleveland) Inc
Reason for recall
A program software error in the algorithm produces image artifacts that could possibly lead to a misdiagnosis of patients who have small metal implants.
Action taken
A 'Field Safety Notice - Urgent - Medical Device Correction' notification was sent to the firm's customers via Certified mail on 3/8/2011. The notification letter explains the nature of the Metal Artifact Reduction/Metal Artifact Correction software error; the Brilliance CT X-Ray systems subject to correction and how to identify them; and the circumstances under which this error can occur. It also asks that the user stop running the MAR application on any of the affected Brilliance systems identified with the recalled serial numbers and software versions alluded to in the medical device correction notification. Lastly, the notification states that Philips Medical will provide a solution that will fix the issue and the notification lists contact names; telephone numbers; and E-mail addresses which the customer can contact for further information concerning this device issue, or customer support assistance.
Root cause
Other
Status
Terminated
Product code
JAK
Quantity affected
10 units
Distribution
Nationwide Distribution -- DC, IL, IN, MI, MO, NY, OH, OR, PA, TX, and VT.
Date initiated
2011-03-07
Date posted
2011-05-19
Date terminated
2013-02-22
Location
Cleveland, OH