Brilliance 64 Computed Tomography System, Model #78231, Serial Number: 9592. The Brilliance 64 is a Computed Tomography

FDA device recall Z-3281-2011 · posted 2011-09-27 · Terminated

Recall details

Recalling firm
Philips Medical Systems (Cleveland) Inc
Reason for recall
Component defect. Philips Medical decided to conduct a recall field /correction (rework) of the mounting key locking threaded insert of the Detector Module system (DMS) bolt on one (1) Brilliance 64 CT X-ray unit. The firm initiated this action following Philip's receipt of the customer's complaint reporting that the retaining inserts were pulling out of the rotor during the removal of the UDMS
Action taken
Philips sent a Field Change Order (FCO) 72800519 on June 13, 2011 to the affected customer. A Field Service Engineer visited the customer site and installed a hardware upgrade to the affected Brilliance 64 device system (Serial Number: 9592). The correction action was completed on August 9, 2011. For questions regarding this recall call 1-440-483-7000.
Root cause
Other
Status
Terminated
Product code
JAK
Quantity affected
1 unit
Distribution
Nationwide Distribution including PA
Date initiated
2011-06-02
Date posted
2011-09-27
Date terminated
2013-03-15
Location
Cleveland, OH