Brilliance 6, 10, 16, 40, 64, Big Bore and the iCT, iCT SP, Ingenuity CT and MX8000 Dual v. EXP are Computed Tomography
FDA device recall Z-0945-2013 · posted 2013-03-13 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Reason for recall
- Philips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.
- Action taken
- Philips Healthcare sent a "Urgent Medical Device Correction" letter dated December 20, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were informed Philips received reports from the field stating that the screws holding the vertical brake together failed. A Field Service Engineer will contact each site to schedule a time to implement the Field Change Order. If you need any further information please contact your local Philips representative or local Philips Healthcare office. For North America and Canada Contact the Customer Care Solutions Center at ( 1-800-722-9377, option 5: Enter Site ID or follow the prompts.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- JAK
- Quantity affected
- 4797 Units
- Distribution
- Worldwide Distribution - Nationwide Distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ,
- Date initiated
- 2012-12-20
- Date posted
- 2013-03-13
- Date terminated
- 2014-04-17
- Location
- Cleveland, OH
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| BRILLIANCE VOLUME (K060937) | Philips Medical Systems (Cleveland), Inc. | 2006-06-05 |
| ACQSIM-MULTISLICE-CT, MODEL 453567042271 (K033357) | Philips Medical Systems (Cleveland), Inc. | 2004-01-09 |
| PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM (K033326) | Philips Medical Systems, Inc. | 2003-10-29 |
| MX8000 IDT CT SYSTEM (K012009) | Marconi Medical Systems, Inc. | 2001-09-13 |
| MX8000 V5.0 CT SYSTEM (K010817) | Philips Medical Systems (Cleveland), Inc. | 2001-06-15 |