Braun Thermoscan PRO 6000 Ear Thermometer

FDA device recall Z-2339-2024 · posted 2024-07-11 · Open, Classified

Recall details

Recalling firm
Baxter Healthcare Corporation
Reason for recall
The Braun Thermoscan PRO 6000 ear thermometers may have been shipped with a compact disc (CD) containing an outdated version of the Instructions for Use (IFU).
Action taken
A communication was sent to affected customers via U.S.P.S., first class mail on 06/13/2024. Baxter is asking customers to: 1. Clinicians may continue to use the PRO 6000 ear thermometers according to the current IFU (CD material Number 421032). Please discard all versions of the outdated IFU (material number 419450). The material number is printed on the CD. 2. Acknowledge the receipt of this notification by completing a reply form on the customer portal. UPDATE PER EMAIL DATED 7/24/2024: The firm issued a follow-up letter dated 7/24/2024 via USPS first class mail. The follow-up letter flagged Urgent Medical Device Correction provides the following additional actions to be taken by the customer as a result of the firm expanding the scope of the recall strategy and actions to be taken by the customer: (The content of the other numbered actions remains the same as the initial letter dated 6/13/2024.) 2. Please share this communication with all potential users in your organization and instruct them to follow cleaning instructions in the IFU Maintenance and Service section for proper cleaning. For convenience, the cleaning instructions are summarized in the enclosed Cleaning Guide. 3. Do not use the device if the ring around the measurement button shows a green blinking or flashing light instead of a ready state (solid green light). Contact Baxter Technical Support to report this issue. 4. Do not use the device if the device requires multiple power-ups prior to going to ready state (solid green light). Contact Baxter Technical Support to report this issue. 5. If you experience an overheating probe tip, do not use the device. Contact Baxter Technical Support to report the issue. A Cleaning Guide was enclosed with the letter as well as another response form.
Root cause
Process control
Status
Open, Classified
Product code
FLL
Quantity affected
265,647 units
Distribution
Distributed to United States, Canada, Australia, Malaysia, New Zealand, Philippines, Austria, Belgium, France, Germany, Italy, Netherlands, Norway, Spain, Sweden, and United Kingdom. UPDATED PER EMAIL
Date initiated
2024-06-13
Date posted
2024-07-11
Location
Deerfield, IL