Brand Name: Welch Allyn, Inc. Product Name: Welch Allyn Spot Vision Screener VS100 Model/Catalog Number: VS100 Software
FDA device recall Z-1573-2026 · posted 2026-03-18 · Open, Classified
Recall details
- Recalling firm
- Baxter Healthcare Corporation
- Reason for recall
- Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.
- Action taken
- On February 18, 2026, the firm notified consignees via Urgent Medical Device Recall letters. Users were instructed to immediately locate, isolate, and stop using the listed affected products. Customers should contact Baxter Technical Support to arrange for the return and correction of their impacted device. Baxter may provide a loaner device upon request while your device is being upgraded and serviced. Distributors should return unopened products in inventory for credit. If affected product was further distributed, the distributor should conduct a consumer-level recall.
- Root cause
- Device Design
- Status
- Open, Classified
- Product code
- HKO
- Quantity affected
- 59
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CO, IL, IN, KS, MA, NE, NJ, NM, NY, PA, SC, TX, VA and the countries of Colombia and India.
- Date initiated
- 2026-02-18
- Date posted
- 2026-03-18
- Location
- Deerfield, IL
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.