Atlas FDA

Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Produc

FDA device recall Z-1491-2026 · posted 2026-03-03 · Open, Classified

Recall details

Recalling firm
Diagnostica Stago, Inc.
Reason for recall
The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.
Action taken
On February 4, 2026 URGENT: MEDICAL DEVICE RECALL (REMOVAL) letters were sent to customers. Actions required: If you currently have one or more kits of STA LIATEST FREE PROTEIN S 6 : - For REF 00516: Lots 271971, 272859, and 273018 with a short expiration date we ask you to discontinue their use and dispose of them in accordance with your local regulations. Stago will provide a credit for any destroyed kits. - For REF 00516: lot 273426: o If the initial barcode sheet provided with the kit has already been loaded into the instrument: - Stop the use the initial barcode pre-calibration settings. - Replace the initial barcode settings with the new pre-calibration barcode settings attached, by following to the procedure described in Appendix 1 of this letter. o If the initial barcode sheet has not yet been loaded into the instrument: - Do not use it and proceed to load the lot using the new pre-calibration barcodes provided in the attachment.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
GGP
Quantity affected
369
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MI, MN, MO, MT, NC, ND, NJ, NY, OH, PA, TN, TX, UT, VA, WA and the countries of ALBANIA, ALGERIA, ARGENTINA, AUS
Date initiated
2026-02-04
Date posted
2026-03-03
Location
Parsippany, NJ