Atlas FDA

Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Pro SuperPulsed Laser System Model/Catalog Numbers:

FDA device recall Z-1715-2025 · posted 2025-05-01 · Open, Classified

Recall details

Recalling firm
Olympus Corporation of the Americas
Reason for recall
Preset treatment parameters are not consistently being used in accordance with the IFU.
Action taken
On April 4, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were sent to customers. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. Olympus will contact you to schedule time for an Olympus Field Representative to visit your facility and install the software update. Additionally, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Please check all areas of your facility to determine if you have the devices specified above. 3. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this notification and the SOLTIVE Laser System Instructions for Use. 4. Olympus requests that you acknowledge receipt of this letter through our recall web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number: "0469 c. Complete the form as instructed. 5. Olympus will contact you to schedule time for an Olympus Field Representative to visit your facility and install the software update. Olympus s record of the completed software update on your SOLTIVE system(s) will serve as the acknowledgment of this field corrective action for your facility. 6. Following the installation of the software update on your SOLTIVE system(s), ensure all personnel are thoroughly trained on the attached IFU Addendum corresponding with this update. The updated version of the full IFU can be located electronically at OlympusConnect.com. If you would like to receive a physical copy of the updated IFU, please submit a request in the comments of our recall portal when submitting your acknowledgement. 7. If you have further distributed this product, identify and forward them this notification. Olympus requests that you report any complaints related to the SOLTIVE Laser System or any associated injuries to our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may al
Root cause
Unknown/Undetermined by firm
Status
Open, Classified
Product code
GEX
Quantity affected
189
Distribution
Worldwide distribution - US Nationwide and the countries of AU, CA, SG, DE, JP, HK.
Date initiated
2025-04-04
Date posted
2025-05-01
Location
Center Valley, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories) (K242191)Olympus Surgical Technologies America2024-08-23