Atlas FDA

Brand Name: NIO-A Product Name: NIO Intraosseous Device Adult Model/Catalog Number: 105000603 Software Version: N/A Prod

FDA device recall Z-2118-2025 · posted 2025-07-17 · Open, Classified

Recall details

Recalling firm
Waismed Ltd.
Reason for recall
Due to manufacturing error, Intraosseous device built-in stabilizer may not be released properly after deployment and insertion.
Action taken
On June 05, 2025, WaisMed issued an "Urgent: Medical Device Recall" notification to affected consignees via E-Mail. WaisMed asked consignees to take the follow actions: 1. Healthcare providers should immediately discontinue use of affected devices and consider alternative intraosseous access products to ensure continuity of patient care without compromising treatment timelines in critical situations. 2. Immediately discontinue, quarantine and remove all NIO A and NIO+ devices from the affected lots from all points of use, including emergency kits, crash carts, ambulances, and clinical storage areas. 3. Verify your inventory against the specific lot numbers listed in this recall notice to identify all affected products. 4. Share and post this Product Recall letter within your facility network. 5. Forward to any customers you may have distributed the product to ensure awareness. 6. Ensure the contents of this Product Recall are read and understood by those within your organization. 7. Please complete the attached Recall Acknowledgement and Receipt Form and return it to: RECALLS@SAFEGUARDMEDICAL.COM or mail to: Waismed Ltd. 10 Amal St. Afek Industrial Park, Rosh Ha'yin 4809234, Israel. Any questions directly associated with this recall should be directed to the: Recall Coordinator or International Director Quality and Regulatory Affairs via email: vigilanceil@safeguardmedical.com, phone: +972-9-9517444 or website: safeguardmedial.com.
Root cause
Process change control
Status
Open, Classified
Product code
FMI
Quantity affected
7056 units (US)
Distribution
Worldwide distribution: US (nationwide) to state of: NC and OUS (international) countries of: Slovenia, Australia, Brazil, Estonia, South Africa, U. Arab Emirate, Greece, Panama, Singapore, Italy, Ger
Date initiated
2025-06-05
Date posted
2025-07-17
Location
Herzliya, Israel

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
NIO -INTRAOSSEOUS DEVICE (K142086)Waismed, Ltd.2014-08-28