Brand Name: MST Duet Reusable Handle for Single Use Heads Product Name: Reusable Handle for Single Use Heads (Forceps an

FDA device recall Z-0521-2025 · posted 2024-11-22 · Open, Classified

Recall details

Recalling firm
MicroSurgical Technology Inc
Reason for recall
Due to pin coming off of handles.
Action taken
On 09/24/2024, the firm sent via express mail a customer letter informing customers that there is a potential issue pertaining to the pins on the handle of the DFH-1028, Handle Assembly Lot Number 160458. The locking pins that actuate the head may be defective and could potentially dislodge without warning. As a precautionary measure, the firm is replacing the handle associated with the affected lot number 160458 at no cost to customers. Customers are being instructed to return the affected products using the supplied return shipping label. Firm's contact information if questions or further assistance is needed: David Wang Sr. VP of Ops QA/RA D: 206-402-1146 8415 154th Avenue NE | Redmond, WA 98052 www.microsurgical.com dwang@microsurgical.com
Root cause
Process control
Status
Open, Classified
Product code
HNF
Quantity affected
24
Distribution
U.S.: CT, GA, MN, MT and TX. O.U.S.: Netherlands 24 total devices = 6 devices distributed in the United States and 18 devices distributed in Germany.
Date initiated
2024-09-17
Date posted
2024-11-22
Location
Redmond, WA