Atlas FDA

Brand Name: Medistim MiraQ System Product Name: Medistim MiraQ System Model/Catalog Number: MQC04001, MQC14001, MQU14011

FDA device recall Z-0410-2026 · posted 2025-10-29 · Open, Classified

Recall details

Recalling firm
Medistim Asa
Reason for recall
Printer Support Module (PSM) may be subject to insufficient soldering, which may cause loosening of the wires and system to shut down. Affected devices need to have their Printer Support Modules replaced.
Action taken
Firm notified impacted consignees by email on September 26, 2025. Consignees were informed that qualified service personnel will perform the correction of the device according to the standard service procedure.
Root cause
Process control
Status
Open, Classified
Product code
DPW
Quantity affected
14
Distribution
Domestic distribution to CO, MI, MN, TN, TX. International distribution to Australia, Austria, Germany, Finland, India, Japan, Russia, China, Saudi Arabia.
Date initiated
2025-09-26
Date posted
2025-10-29
Location
Oslo, Norway