Atlas FDA

Brand Name: HeartMate Product Name: HeartMate 3" System Controller Model/Catalog Number: 106531US Software Version: N/A

FDA device recall Z-2519-2026 · posted 2026-06-18 · Open, Classified

Recall details

Recalling firm
Thoratec LLC
Reason for recall
Due to 11 volt Backup Battery failures.
Action taken
On 05/13/2026, the firm sent email an "URGENT MEDICAL DEVICE RECALL" Letter informing customers that As of April 27, 2026, Abbott received twenty-four (24) complaints where an advisory warning, Replace Backup Battery, was triggered once certain 11V Backup Batteries were connected to the System Controller. This warning is displayed as a yellow banner on the HeartMate Touch Monitor and System Monitor. This advisory alert is only visible when using HeartMate Touch or the System Monitor in the clinic, it is not visible to the patient. Customers are instructed to: 1. Read this notification thoroughly, complete and sign the form included with this letter, acknowledging that you received and understand this information, and that you will communicate this information to those relevant in your facility. Once complete, return the acknowledgement form to Abbott. 2. Use the online lookup tool (https://www.cardiovascular.abbott/us/en/hcp/product-advisories/heartmatebatteries-2026.html) to confirm if a 11V Lithium-Ion battery at your site is an impacted one. Figure 3 shows an instruction on how to locate the serial numbers of the impacted batteries. 3. Once the impacted batteries are identified, do not use them. 4. Beginning May 15, 2026, a separate email will be sent to your site confirming shipment of replacement batteries. This email will include return shipping labels and a device retrieval form, along with detailed instructions for returning the affected batteries. For Questions - contact your Abbott representative or Abbott Technical Support at 1-800-456-1477 (Business hours: 8 AM EST to 7 PM EST).
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
DSQ
Quantity affected
694
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, DC, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA
Date initiated
2026-05-13
Date posted
2026-06-18
Location
Pleasanton, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.